The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateDecember 08, 2025
ReportedJanuary 28, 2026
Recalling firmBeckman Coulter Ireland, Inc.
Country / regionIreland US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of

Hazard

Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).

Product description

The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026