Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | August 28, 2025 |
| Reported | November 05, 2025 |
| Recalling firm | Vantive US Healthcare LLC |
| Country / region | United States IL |
Hazard
Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.
Affected products
- Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (
Related recalls
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER