HomeMedical Device recalls
Medium severity

Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateAugust 28, 2025
ReportedNovember 05, 2025
Recalling firmVantive US Healthcare LLC
Country / regionUnited States IL

Hazard

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026