HomeMedical Device recalls
Medium severity

Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateDecember 11, 2025
ReportedFebruary 11, 2026
Recalling firmFujirebio Diagnostics, Inc.
Country / regionUnited States PA

Hazard

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Product description

Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology13 by a specific two-step immunoassay method on the LUMIPULSE G System. Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-¿-amyloid 1-42 monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-¿-amyloid 1-42 monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains detergents in 50 mM MES buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026