Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 7222
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | March 31, 2026 |
| Reported | May 20, 2026 |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Country / region | Netherlands Worldwide distribution - US Nationwide and the country of Canada. |
Hazard
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
Product description
Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).
Affected products
- Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (7222
Related recalls
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER