Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateOctober 02, 2025
ReportedDecember 03, 2025
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Country / regionNetherlands N/A

Hazard

The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026