HomeMedical Device recalls
High severity

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE

SourceFDA Device Enforcement
CategoryMedical Device
SeverityHigh
Statusongoing
Recall dateFebruary 06, 2026
ReportedMarch 18, 2026
Recalling firmTrividia Health, Inc.
Country / regionUnited States FL

Hazard

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Product description

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i81-11BK, TRUE METRIX GO Kit (United Kingdom - mmol/L)/RF4i82-11BK, TRUE METRIX GO Starter Kit (United Kingdom - mmol/L)/RF4i82-12BK

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026