HomeMedical Device recalls
Medium severity

Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateFebruary 27, 2026
ReportedApril 15, 2026
Recalling firmSpacelabs Healthcare, Inc.
Country / regionUnited States WA

Hazard

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026