HomeMedical Device recalls
Medium severity

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateDecember 10, 2025
ReportedJanuary 21, 2026
Recalling firmElekta, Inc.
Country / regionUnited States GA

Hazard

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026