Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Nu
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | January 07, 2026 |
| Reported | February 25, 2026 |
| Recalling firm | Medline Industries, LP |
| Country / region | United States IL |
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Product description
Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 15) NEURO PACK, Model Number: DYNJ49098I; 16) NEURO PACK, Model Number: DYNJ53089B; 17) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 18) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 19) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 20) NEURO PACK, Model Number: DYNJ60792B; 21) NEURO PACK, Model Number: DYNJ60792C; 22) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 24) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 25) NEURO PACK, Model Number: DYNJ61740D; 26) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 27) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 28) DISCECTOMY PACK, Model Number: DYNJ67835; 29) WWD NEURO PACK, Model Number: DYNJ80849B; 30) WWD NEURO PACK, Model Number: DYNJ80849J; 31) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 32) NEURO, Model Number: DYNJ900904J; 33) NEURO, Model Number: DYNJ900904K; 34) NEURO, Model Number: DYNJ900904L; 35) NEURO, Model Number: DYNJ900904M; 36) NEURO, Model Number: DYNJ902585; 37) NEURO, Model Number: DYNJ902585A; 38) NEURO - IFCAP 230316, Model Number: DYNJ908393; 39) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 40) NEURO KIT, Model Number: DYNJ909573; 41) NEURO, Model Number: DYNJ910384; 42) NEURO, Model Number: DYNJ910384A; 43) NEURO PACK-LF, Model Number: PHS390015M
Affected products
- Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRAN