Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number: DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number: DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number: DYNDH1934;
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | January 07, 2026 |
| Reported | February 25, 2026 |
| Recalling firm | Medline Industries, LP |
| Country / region | United States IL |
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Product description
Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number: DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number: DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number: DYNDH1934; 6) BREAST BIOPSY PACK, Model Number: DYNJ17489I; 7) VOR PACK-LF, Model Number: DYNJ48539; 8) VOR PACK-LF, Model Number: DYNJ48539A; 9) IR BIOPSY PACK, Model Number: DYNJ59897B; 10) BREAST PACK, Model Number: DYNJ82998
Affected products
- Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIO