MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | High |
| Status | ongoing |
| Recall date | November 26, 2025 |
| Reported | January 28, 2026 |
| Recalling firm | Medline Industries, LP |
| Country / region | United States IL |
Hazard
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Affected products
- MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
Related recalls
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER