HomeMedical Device recalls
Medium severity

MC3 VitalFlow Console, REF 58100; Blood pump of ecmo

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateOctober 03, 2025
ReportedNovember 26, 2025
Recalling firmMedtronic Perfusion Systems
Country / regionUnited States MN

Hazard

As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026