MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateDecember 10, 2025
ReportedJanuary 21, 2026
Recalling firmREPLIGEN CORPORATION
Country / regionUnited States MA

Hazard

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026