IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | High |
| Status | ongoing |
| Recall date | September 30, 2025 |
| Reported | November 05, 2025 |
| Recalling firm | IMPRIS Imaging Inc |
| Country / region | United States MN |
Hazard
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Affected products
- IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
Related recalls
Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755PMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass vMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass vaSwan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros;