HomeMedical Device recalls
Medium severity

ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateDecember 15, 2025
ReportedJanuary 28, 2026
Recalling firmICU Medical, Inc.
Country / regionUnited States IL

Hazard

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026