I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2.
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | January 23, 2026 |
| Reported | April 01, 2026 |
| Recalling firm | I.T.S. GmbH |
| Country / region | Austria US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W. |
Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Product description
I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.
Affected products
- I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (D
Related recalls
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER