HomeMedical Device recalls
Medium severity

I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2.

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateJanuary 23, 2026
ReportedApril 01, 2026
Recalling firmI.T.S. GmbH
Country / regionAustria US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Hazard

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Product description

I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026