i-STAT G3+ cartridge; List Number: 03P78-26;
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | April 01, 2026 |
| Reported | May 06, 2026 |
| Recalling firm | Abbott Point Of Care Inc. |
| Country / region | United States NJ |
Hazard
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
Affected products
- i-STAT G3+ cartridge; List Number: 03P78-26;
Related recalls
Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755PMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass vMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass vaSwan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros;