ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | March 31, 2026 |
| Reported | May 13, 2026 |
| Recalling firm | Erbe Medical, LLC |
| Country / region | United States AZ |
Hazard
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Affected products
- ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;
Related recalls
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER