DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Module, DxC 500i (Part Number C13252)
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | September 19, 2025 |
| Reported | October 29, 2025 |
| Recalling firm | Beckman Coulter, Inc. |
| Country / region | United States MN |
Hazard
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
Affected products
- DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Modul
Related recalls
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp.Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER