Medium severity

CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateOctober 10, 2025
ReportedJanuary 21, 2026
Recalling firmThoratec LLC
Country / regionUnited States CA

Hazard

Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.

Product description

CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026