CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | October 10, 2025 |
| Reported | January 21, 2026 |
| Recalling firm | Thoratec LLC |
| Country / region | United States CA |
Hazard
Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.
Product description
CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System
Affected products
- CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMa