HomeMedical Device recalls
Medium severity

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusabl

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateOctober 08, 2025
ReportedNovember 19, 2025
Recalling firmOlympus Corporation of the Americas
Country / regionUnited States PA

Hazard

The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.

Product description

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026