HomeMedical Device recalls
Medium severity

Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description:

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateMarch 23, 2026
ReportedMay 06, 2026
Recalling firmAdvanced Bionics, LLC
Country / regionUnited States CA

Hazard

The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.

Product description

Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280 Software Version: Not Applicable Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida" CI M and Sky CI" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implant battery (P675). Regular 675 hearing instrument batteries will not be powerful enough for a cochlear implant. The suggested manufacturer and battery is PowerOne P675 Implant Plus battery. The M Zn-Air Battery Pak is available with non-tamperproof and tamperproof cartridges. Component: Not Applicable

Affected products

Related recalls

3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS- · May 20, 2026QuickVue Dipstick Strep A Test, 50T, REF 20108, antigens, all groups, streptococcus spp. · May 20, 2026Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Number: 11-1424 Software Version: · May 20, 2026Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25 · May 20, 2026BW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER · May 20, 2026