BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted route
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | November 06, 2025 |
| Reported | February 04, 2026 |
| Recalling firm | CareFusion 303, Inc. |
| Country / region | United States CA |
Hazard
Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network
Product description
BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. The BD AlarisTM System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, patient-controlled analgesia (PCA) Module, and associated software applications.
Affected products
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring sys