Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079, 722224).
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | April 28, 2026 |
| Reported | May 27, 2026 |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Country / region | Netherlands Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen; |
Hazard
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
Affected products
- Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion
Related recalls
Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755PMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass vMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass vaSwan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros;