Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIR
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | High |
| Status | ongoing |
| Recall date | October 10, 2025 |
| Reported | November 19, 2025 |
| Recalling firm | Medline Industries, LP |
| Country / region | United States IL |
Hazard
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Product description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;
Affected products
- Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA1