HomeMedical Device recalls
Medium severity

Allura Xper FD10; Model Number: 722026;

SourceFDA Device Enforcement
CategoryMedical Device
SeverityMedium
Statusongoing
Recall dateDecember 15, 2025
ReportedJanuary 21, 2026
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Country / regionNetherlands Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.

Hazard

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Affected products

Related recalls

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94215T; Cardiopulmonary bypass v · May 27, 2026Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94665; Cardiopulmonary bypass va · May 27, 2026Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, · May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; · May 27, 2026