(1)(2) LeadCare¿ II Blood Lead Test Kit, (3)LeadCare¿ II 5X Blood Lead Test Kit, (4)LeadCare¿ II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 70-6762RUO, (3) 70-8404 (Box containing 5 Test Kits), (4) 70-9000 (Box containing 9 test kits); The LeadCare II kit is provided with the fol
| Source | FDA Device Enforcement |
|---|---|
| Category | Medical Device |
| Severity | Medium |
| Status | ongoing |
| Recall date | March 13, 2026 |
| Reported | April 29, 2026 |
| Recalling firm | Magellan Diagnostics, Inc. |
| Country / region | United States MA |
Hazard
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Product description
(1)(2) LeadCare¿ II Blood Lead Test Kit, (3)LeadCare¿ II 5X Blood Lead Test Kit, (4)LeadCare¿ II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 70-6762RUO, (3) 70-8404 (Box containing 5 Test Kits), (4) 70-9000 (Box containing 9 test kits); The LeadCare II kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50 @) " Transfer Droppers (50 @) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Affected products
- (1)(2) LeadCare¿ II Blood Lead Test Kit, (3)LeadCare¿ II 5X Blood Lead Test Kit, (4)LeadCare¿ II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 7